1.
Chinese Journal of Industrial Hygiene and Occupational Diseases ; (12): 932-935, 2019.
Article in Chinese | WPRIM | ID: wpr-799921

ABSTRACT

Objective@#Improvement the method of nitrogen oxide sampling, analysis and detection in workplace air.@*Methods@#Absorption and collection with sodium carbonate-sodium bicarbonate buffer solution, analysised by ion chromatography, anion column AS23, column temperature 30 ℃.@*Results@#The linearity of the nitrite ion in the corresponding mass concentration range 0~90.0 μg/ml was good (Nitrogen dioxide in 0~120.0 mg/m3) , correlation coefficient>0.999 5. The detection limit was 0.16 μg/ml, The minimum detection concentration was 0.11 mg/m3 with the sampling volume 7.5 L.@*Conclusion@#The method has good accuracy, high sensitivity, with directing injection and no pre-treatment, and can popularize in general laboratories, and can satisfy the determination of nitrogen oxides in the air of workplace.

2.
Chinese Journal of Industrial Hygiene and Occupational Diseases ; (12): 633-636, 2018.
Article in Chinese | WPRIM | ID: wpr-807060

ABSTRACT

Objective@#To assess psychological acceptance and occupational stress of medical staff, analyze the relationship among personality, psychological acceptance and occupational stress and discuss the direct or indirect effects of personality to occupational stress.@*Methods@#The gaseous four kinds of carboxylic acids in the workplace air were simultaneously collected by silica gel tube, and then desorbed by deionized water and eluted by ion chromatograph. The the content was detected by conductivity detector.@*Results@#The linear relationship was good in the concentration range of about 0~140 mg/L. The correlation coefficient r>0.999, and the maximum detection limit was 4.5 μg/mL. The sampling efficiency of the four carboxylic acids ranges from 96.10%~100.27%. Through the sample added recovery experiments, the low and high content of the silicone tube samples were detected; and the range of desorption efficiency was 82.18%~100.12%; the range of precision was 0.70%~3.71%.@*Conclusion@#This method adopts deionized water to desorb samples, and the application of ion chromatography detection have reached the requirements of《Guidelines for the establishment of occupational health standards, Part 4: Determination of chemical substances in workplace air》, which can be used in four kinds of workplace air detection of carboxylic acid compounds.

3.
Chinese Journal of Industrial Hygiene and Occupational Diseases ; (12): 468-471, 2018.
Article in Chinese | WPRIM | ID: wpr-806624

ABSTRACT

Objective@#To establish an HPLC method for determining p-Anisidine in workplace air.@*Methods@#An C18 capillary column was used with the detection wavelength of 240 nm, the methanol solution (v/v=62:38) as mobile phase, the flow rate was 0.9 ml/min, and the injection volume was 20 μl.@*Results@#The linear range (μg/ml) for p-Anisidine was 0~100; correlation coefficient was 0.999 7; the detection limit was 0.15 μg/ml; the average precision of RSD was 4.4%. The sampling efficiency was more than 95% using silicone tube as sampling filter. Sample can be stored at room temperature for more than 7 d.@*Conclusion@#the method appered to be accurate, simple, sensitive and reliable, also have a high sampling efficiency, and can be usd to determine p-anisidine in workplace air simultaneously.

4.
Chinese Journal of Industrial Hygiene and Occupational Diseases ; (12): 474-476, 2015.
Article in Chinese | WPRIM | ID: wpr-350567

ABSTRACT

<p><b>OBJECTIVE</b>To establish a rapid determination method with gas chromatography-mass spectrometer (GC-MS) for thiodiglycolic acid (TDGA), a vinyl chloride (VCM) biomarker.</p><p><b>METHODS</b>A high- sensitivity determination method was established using a moderate methyl esterification instead of methyl esterification of highly toxic diazo reaction.</p><p><b>RESULTS</b>The standard curve regression linear equation of the method was: y=8460.5x-4758.2, the linear coefficient was 0.999 7, the minimum quantity concentration was 2.0 µg/L, the range of precision value was 0.81%-2.38%, and the average recovery of standard addition was 99.0%-102.9%.</p><p><b>CONCLUSION</b>This method reduces the risk of traditional methyl esterification, improves the determination sensitivity compared with the GC-FPD method, and meets the determination requirement of TDGA.</p>


Subject(s)
Humans , Gas Chromatography-Mass Spectrometry , Thioglycolates , Urine , Vinyl Chloride